职业与健康 ›› 2025, Vol. 41 ›› Issue (3): 341-344.

• 论著 • 上一篇    下一篇

便携式离子阱质谱技术同时测定工作场所中10种常见有机毒物

庄小舟, 蔡小璇, 李思燕, 陈满连, 蔡木蔚   

  1. 东莞市第六人民医院(东莞市职业病防治中心)职业健康监护与监测科,广东 东莞 523008
  • 收稿日期:2024-06-01 修回日期:2024-06-11 出版日期:2025-02-01 发布日期:2025-12-17
  • 作者简介:庄小舟,男,副主任技师,主要从事职业卫生监测与评价工作。
  • 基金资助:
    东莞市社会发展科技面上项目(20211800900962); 东莞市社会发展科技重点项目(20231800935812); 广东省医学科研基金项目(B2022254)

Simultaneous detection of 10 kinds of common organic toxicants in workplace by a portable ion trap mass spectrometer technique

ZHUANG Xiaozhou, CAI Xiaoxuan, LI Siyan, CHEN Manlian, CAI Muwei   

  1. Occupational Health Surveillance and Monitoring Department,Dongguan Sixth People's Hospital(Dongguan Center for Occupational Diseases Control and Prevention),Dongguan,Guangdong 523008,China
  • Received:2024-06-01 Revised:2024-06-11 Online:2025-02-01 Published:2025-12-17

摘要: 目的 建立便携式离子阱质谱技术同时测定工作场所中10种常见有机毒物的方法。方法 采用仪器探头进行采样和检测,DB-5MS色谱柱分离,以保留时间和离子阱质谱特征离子定性,标准曲线法定量。结果 自建离子阱质谱库比NIST质谱库定性结果更可靠。常见有机毒物在2.00~50.0 ppb呈线性关系,线性系数均>0.999 0;方法最低检出浓度范围为0.05~0.31 μg/m3;最低定量浓度范围为0.16~1.02 μg/m3。检测结果批间精密度为1.6%~5.4%(n=6),批内精密度为1.0%~4.1%(n=6),加标回收率为95.3%~101.3%。工作场所多种有机毒物不干扰测定。结论 所建立的方法适用于工作场所常见有机毒物同时测定,方法灵敏度和最低检出限优于便携式四级杆质谱法。

关键词: 便携式, 离子阱, 质谱器, 有机毒物, 方法研究

Abstract: Objective To establish a method for simultaneously detecting ten kinds of common organic toxicants in workplace by a portable ion trap mass spectrometer technique. Methods The instrument probe was used for sampling and detection. The samples were separated with DB-5MS chromatographic column. The retention time and characteristics of the ions by ion trap mass spectrometry were used for qualitative analysis,and standard curve method was used for quantitative analysis. Results The qualitative results of the self-built ion trap mass spectrometer library were more reliable than those of the NIST mass spectrometer library. The determination of common organic toxicants were linear relation in the concentration range of 2.00-50.0 ppb,and the linear correlation coefficients were all >0.999 0. The minimum detection concentration was 0.05-0.30 μg/m3,and the minimum quantitation concentration was 0.16-1.02 μg/m3. The within-run and the between-run relative standard deviation were 1.6%-5.4%(n=6) and 1.0%-4.1%(n=6) respectively,and the standard recovery rate was 95.3-101.3%. Multiple organic toxicants in the workplace had no interference with the measurement results. Conclusion This method is suitable for simultaneous determination of common organic toxicants in workplace,and the sensitivity and minimum detection limit of this method are superior to that of portable quadrupole mass spectrometry.

Key words: Portable, Ion trap, Mass spectrometry, Organic toxicants, Methods research

中图分类号: