摘要:

目的 建立全自动固相微萃取-气相色谱-三重四极杆质谱法测定水中环氧氯丙烷和亚硝基二甲胺的方法。方法 10 mL水样加入4 g氯化钠后直接测定,采用二乙烯基苯/炭分子筛/聚二甲基硅氧烷(divinylbenzene/carbon wide range/polydimethylsiloxane,DVB/C-WR/PDMS)纤维在30 ℃、500 rpm搅拌下萃取15 min,于250 ℃解析1 min。气相色谱不分流进样,采用InertCap Pure-Wax(30 m×0.25 mm×0.25 μm)色谱柱程序升温分离,质谱多反应监测(multiple reaction monitoring,MRM)模式采集数据,外标法定量。结果 环氧氯丙烷于0.02~0.80 μg/L、亚硝基二甲胺于0.05~2.00 μg/L范围内均线性良好,相关系数>0.999 3,方法检出限环氧氯丙烷为0.008 μg/L,亚硝基二甲胺为0.02 μg/L,低、中、高浓度平均加标回收率为89.1%~104.0%,相对标准偏差为6.1%~9.7%。结论 本方法适用于水中环氧氯丙烷和亚硝基二甲胺的检测,方法灵敏度满足各类水质监测标准要求,便捷高效、环境友好。

关键词: 水, 固相微萃取, 环氧氯丙烷, 亚硝基二甲胺, 气相色谱-三重四极杆质谱, 绿色化学

Abstract:

Objective To establish an auto solid phase microextraction-gas chromatography-triple quadrupole mass spectrometry to detect epichlorohydrin(ECH) and n-nitrosodimethylamine(NDMA) in water. Methods Water sample was used 10 mL and added 4 g NaCl. The optimal extraction conditions were achieved with SPME fiber of divinylbenzene/carbon wide range/polydimethylsiloxane(DVB/C-WR/PDMS). Sample was extracted at 30 ℃ and 500 rpm for 15 min,and desorbed at 250 ℃ for 1 min. Chromatographic separation was performed on a InertCap Pure-Wax chromatographic column(30 m×0.25 mm×0.25 μm),and splitless injection. The target compounds were detected by the multi reaction monitoring(MRM) mode. The quantitation was carried out by the external standard method. Results ECH showed the good linearities in 0.02-0.80 μg/L,NDMA showed the good linearities in 0.05-2.00 μg/L, and the correlation coefficients(r) were >0.999 3. The limits of detection(LODs,S/N=3) for ECH and NDMA were 0.008 and 0.02 μg/L. The average recoveries ranged from 89.1% to 104.0%,and relative standard deviation(RSD) were in the range of 6.1%-9.7% with the concentration levels of low,medium and high. Conclusion This method is applicable for the detection of ECH and NDMA in water,and the method sensitivity meets the requirements of various water quality monitoring standards,which is convenient, efficient,and environmentally friendly.

Key words: Water, Solid phase microextraction, Epichlorohydrin, N-nitrosodimethylamine, Gas chromatography-triple quadrupole mass spectrometry, Environmentally benign chemistry