职业与健康 ›› 2025, Vol. 41 ›› Issue (6): 736-740.

• 论著 • 上一篇    下一篇

乌鲁木齐市某医院OUR-XGD型伽玛刀对周围环境辐射水平影响和工作人员个人剂量评估

李文哲, 徐丽, 夏宏娟, 刘陈晨, 黄裔玮, 袁芳   

  1. 新疆军区总医院放疗科,新疆 乌鲁木齐 830000
  • 收稿日期:2024-10-21 修回日期:2024-11-18 出版日期:2025-03-15 发布日期:2025-12-16
  • 通信作者: 袁芳,主任医师,E-mail:zyyfmxj@163.com
  • 作者简介:李文哲,男,主管技师,主要从事肿瘤放疗治疗与质控工作。
  • 基金资助:
    新疆维吾尔自治区自然科学基金(2021D01C471)

Impact of OUR-XGD gamma knife on environmental radiation levels and personal radiation dose assessment of staff in a hospital in Urumqi City

LI Wenzhe, XU Li, XIA Hongjuan, LIU Chenchen, HUANG Yiwei, YUAN Fang   

  1. Department of Radiotherapy,General Hospital of Xinjiang Military Command,Urumqi,Xinjiang 830000,China
  • Received:2024-10-21 Revised:2024-11-18 Online:2025-03-15 Published:2025-12-16
  • Contact: YUAN Fang,Chief physician,E-mail:zyyfmxj@163.com

摘要: 目的 研究头部伽玛刀装置周围主要环境关注点辐射水平,并对在此环境中工作的医务人员个人剂量进行评估。方法 利用电离室剂量计测量2011年11月—2021年4月伽玛刀焦点剂量率,同时利用SW88型X-γ辐射剂量仪对OUR-XGD型头部伽玛刀治疗机房内外不同环境关注点辐射水平进行监测。以靠近治疗床辅助患者治疗摆位时站立位置的周围剂量当量率为计算依据,在治疗一定数量患者情况下,近似计算得到工作人员个人剂量。结果 伽玛刀焦点剂量率呈指数形式衰减,伽玛刀周边环境关注点辐射水平监测显示,工作人员上机摆位治疗时站立位置辐射水平由初装时5.100 μSv/h衰减为1.145 μSv/h,伽玛刀机房边缘处辐射水平已降低至接近本底辐射水平。通过近似计算,伽玛刀按年均治疗200人次计算,每人次医务人员驻留机房时间10 min,计算得到治疗期间个人剂量总和约为0.094 mSv,满足电离辐射防护与辐射源安全标准。结论 正常治疗条件下OUR-XGD型奥沃头部伽玛刀操作人员在辅助患者摆位过程中接受的γ辐射满足国家标准,且在屏蔽良好情况下对治疗室内外环境辐射水平无明显影响。

关键词: 伽玛刀, 放射源, 个人剂量, 焦点剂量率, 剂量当量率

Abstract: Objective To explore the radiation levels of the main environmental concerns around the head Gamma Knife device and evaluate the personal doses of medical personnel working in this environment. Methods The ionization chamber dosimeter was used to detect focal dose rate of gamma knife from November 2011 to April 2021. At the same time,the SW88 X-γ radiation dosimeter was used to monitor the radiation levels of different environmental focus points inside and outside the OUR-XGD head gamma knife treatment room. Based on dose equivalent rate of the standing position where the medical staff stood close to the treatment bed for assisted treatment and a certain number of patients,personal dose of the staff member can be obtained by approximate calculation. Results Focal dose rate of gamma knife decayed exponentially.The radiation level monitoring of the surrounding environment concerns of the gamma knife showed that during the staff-assisted positioning treatment the average dose equivalent rate in the standing position decreased from 5.10 μSv/h to 1.145 μSv/h. Furthermore,the dose equivalent rate at the edge of the Gamma Knife machine room had been reduced to near background radiation levels. Based on the average annual treatment of 200 people and mean retention time of 10 minutes for each patient,total personal dose was approximately calculated with the value of 0.094 mSv,which met the standards for ionizing radiation protection and radiation source safety. Conclusion Under normal treatment conditions,the personal dose level of operating staff from OUR-XGD gamma knife during the process of assisting patient positioning meets the safety requirements of national standards. Moreover,it has no obvious effect on the radiation level of indoor and outdoor environment under the condition of good shielding.

Key words: Gamma knife, Radioactive source, Personal dose, Focal dose rate, Dose equivalent rate

中图分类号: