加速器典型照射条件下靶外组织受照剂量测量与分析

姜雯雯1,2, 顾浩琰1,2, 于润秋2, 冯泽臣2, 王宏芳2, 马永忠1,2()   

  1. 1.首都医科大学,北京 100069
    2.北京市疾病预防控制中心,北京 100013
  • 收稿日期:2025-11-28 修回日期:2025-12-26 出版日期:2026-01-01 发布日期:2026-02-03
  • 通信作者: 马永忠,E-mail:myz0905@126.com
  • 作者简介:姜雯雯,女,在读硕士研究生,研究方向为放射防护、辐射监测与健康风险。
  • 基金资助:
    首都卫生发展科研专项项目(首发2024-2G-30119);北京市高层次公共卫生技术人才建设项目(学科带头人-01-20)

Measurement and analysis of out-of-field tissue dose under typical accelerator radiotherapy conditions

JIANG Wenwen1,2, GU Haoyan1,2, YU Runqiu2, FENG Zechen2, WANG Hongfang2, MA Yongzhong1,2()   

  1. 1. Capital Medical University,Beijing 100069 China
    2. Beijing Center for Disease Prevention and Control,Beijing 100013,China
  • Received:2025-11-28 Revised:2025-12-26 Online:2026-01-01 Published:2026-02-03

摘要:

目的 掌握加速器典型放射治疗条件下患者照射野靶外组织受照剂量水平,为患者防护方案的优化和精确治疗计划的制定与实施提供实验依据。方法 以Varian TrueBeam 6 MV直线加速器为辐射源,采用Phantom Laboratory公司生产的RANDO 110成人女性仿真模体,在10 cm×10 cm标准照射野、SSD 100 cm条件下对脐下腹部区域累积照射600 cGy,用LiF(Mg、Cu、P)热释光探测器测量模体乳腺、肝脏、胃、膀胱及性腺等关键组织器官受照剂量。采用Excel 2019建立数据库,SPSS 21.0软件进行统计分析,Mann-Whitney检验(U检验)比较照射野左右两侧人体躯干从胸部到盆部的纵向剂量水平。结果 照射野靶外邻近器官受照剂量最高,其中左右侧性腺平均剂量为723.00 mGy(为照射野中心剂量的12.05%),膀胱平均剂量为409.00 mGy(6.82%),肝脏剂量为46.40 mGy(0.77%)。照射野靶外远端器官受照剂量相对较低,左右侧乳腺平均剂量分别为10.20 mGy(0.17%)和11.01 mGy(0.18%)。照射野左右两侧的剂量分布差异无统计学意义(Z=1.132,P>0.05)。结论 放射治疗导致的靶外组织受照不可避免但不容忽视,照射野靶外邻近组织器官的受照剂量甚至可能达到放射损伤阈值,临床实践中应采取个性化防护策略,最大限度降低靶外受照剂量与辐射风险。

关键词: 放射治疗, 靶外组织, 患者剂量, 放射防护, 仿真模体, 直线加速器

Abstract:

Objective To investigate the dose of out-of-field tissues under typical linear accelerator radiotherapy conditions,and to provide experimental evidence for optimizing patient protection strategies and refining treatment planning. Methods A Varian TrueBeam 6 MV linear accelerator served as the radiation source,and an adult female anthropomorphic phantom RANDO 110 manufactured by Phantom Laboratory was used to simulate the patient in this study. A 10 cm×10 cm standard field was applied to the abdomen below navel with SSD of 100 cm,delivering a total dose of 600 cGy. LiF(Mg,Cu,P) thermoluminescent detectors were used to measure the absorbed doses in critical out-of-field tissues,including the breast,liver,stomach,bladder and gonads. A database was established using Excel 2019,and SPSS 21.0 software was utilized for statistical analysis. The Mann-Whitney test(U test) was employed to compare the longitudinal dose levels from the chest to the pelvis in the human torso on both the left and right sides of the radiation field. Results The highest out-of-field doses were observed in organs adjacent to the radiation field. The mean doses to the left and right gonads were 723.00 mGy(12.05% of the target dose),and the mean bladder dose was 409.00 mGy(6.82%). The liver received a dose of 46.40 mGy(0.77%). In contrast,doses to remote organs were considerably lower than that of above,with mean left and right breast doses of 10.20 mGy(0.17%) and 11.01 mGy(0.18%),respectively. There was no statistically significant difference in the distribution between the left and right sides dose(Z=1.132,P>0.05). Conclusion Radiation exposure of out-of-field tissues during radiotherapy is inevitable and non-negligible. Doses to organs close to the target-field may approach thresholds for radiation-induced damage. Therefore,individualized protection strategies should be implemented in clinical practice to minimize out-of-field exposure and associated radiation risks.

Key words: Radiotherapy, Out-of-field tissues, Patient absorbed dose, Radiation protection, Anthropomorphic phantom, Linear accelerator